Adjustmentandfunctionalities
A
to calibrate automatically
→
Briefly press push button.
Calibration in progress
End of calibration: Red LED
goes off?
YES
Zero point stored.
Go back to Read mode.
.....
Fig. 27 : Calibration of the zero flow point, part 2
Type 8041
to select the other cali-
bration operation
→
Red LED flashes briefly
NO
twice: zero point has not
been calibrated.
→
Briefly press push button
Go back to Read mode
using the zero point of the
previous calibration.
B
Press and hold the push
button until...
...bargraph goes back to
step 1, Fig. 26
→
Check that the calibration
instructions are respected
(see beginning § 10.7).
C
to go back to Read mode
→
Wait 10 s
Go back to Read mode
without confirming the
selection displayed.
→
Restart calibration
of the zero flow
point.
English
33