115
Fig. 4 :
Dimensions of the 8041 transmitter [mm]
6.3.2.

General data

Pipe diameter
DN06 to DN400
Type of fitting
S020
10
88
M20 x 1.5
English
Type 8041
Technicaldata
Fluid temperature
The fluid temperature may be restricted
by the fluid pressure and the material of
the used S020 fitting (see Fig. 5 and
Fig. 6 ).
Sensor holder in
-15...+150 °C
stainless steel
Sensor holder in PVDF
0...+80 °C
Fluid pressure
The fluid pressure may be restricted by
the fluid temperature and the material of
the used S020 fitting (see Fig. 5 and
Fig. 6 ).
Sensor holder in
PN10 with plastic fitting, PN16 with
stainless steel
metal fitting
Sensor holder in PVDF
PN10
Fluid conductivity
> 20 µS/cm
Measurement range
0,2 m/s to 10 m/s
Measurement accuracy
for value measured from 1 to 10 m/s and
-15°C < Tfluid < +130°C for stainless
steel sensor armature
or 0°C < Tfluid < +80°C for PVDF
sensor armature
with Teach-in
• ±2 % of the measured value
with standard K factor
• ±4 % of the measured value
≤ ±(1 % of the measured value + 0.1%
Linearity
of the full scale)
with full scale = 10 m/s