Automatic Calibration (Aut-Cal) - Adam Equipment Solis SAB Série Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

10.12.1

Automatic calibration (AUT-CAL)

The balance self-calibrates when the temperature variation exceeds the factory pre-set value
and at the factory pre-set time intervals, through the internal reference mass, and only if the
balance pan is empty.
When the balance needs to perform the automatic calibration, the display will show the
following message:
A 25-seconds countdown will start during which you can decide to:
Stop the automatic-calibration procedure by pressing the ON/OFF button that will be
delayed of 5 minutes.
or
Let the countdown finish so that the automatic calibration starts.
Note: during the countdown do NOT load anything on the weighing pan!
In this mode, it is also possible to carry out the calibration with the internal reference mass by
pressing the CAL button at any moment, first ensuring that no weight is loaded on the weighing
pan.
1. Press the CAL button with the pan empty. The display will show the message "CAL" and
the balance's calibration will be carried out automatically.
©Adam Equipment Co. Ltd. 2018
31

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Solis spb série

Table des Matières