4. Remove the calibration weight. The balance is now ready for weighing operation.
NOTE: an error message will be displayed if there is some interference during the
calibration process. To interrupt the calibration process, press the ON/OFF button while
the calibration weight indication flashes.
It is also possible to calibrate the balance with a calibration weight greater than the pre-set
calibration weight:
1. Press and hold the CAL button with the weighing pan empty, until the beeping stops, and
then release the button. The word "-CAL-" will be shown on the display, followed by the
word "LOAD", flashing.
2. Load a weight that is equal to or greater than the pre-set calibration weight onto the
weighing plate; the balance will recognize a weight that is equal to or greater than the
calibration weight as valid provided that it is a recognised weight with respect to the most
significant figure of the calibration weight.
e.g.: if the calibration weight is 200 g, it will be possible to calibrate the balance with values
that go from 200g, 300g, 400g up to the upper capacity limit of the balance. The word
"LOAD" on the display will stop flashing; once the calibration has been carried out, the
value of the weight used will be displayed.
©Adam Equipment Co. Ltd. 2018
15