8.0

CALIBRATION

The electronic balance carries out mass measurements using gravity. Differences in
geographical regions and altitudes vary the gravitational acceleration.
The balance must therefore be adapted to environmental conditions to obtain accurate
measurements. This regulation is carried out through the calibration function.
8.1

EXTERNAL CALIBRATION

The calibration is carried out through pressing the CAL button.
1. Press the CAL button with the plate unloaded; the word CAL will be displayed.
2. When the value of the calibration weight begins to flash, load the weight shown on the
display onto the weighing pan.
3. The display will stop flashing, indicating the value of the calibration weight with the stability
indicator on.
Once the calibration has been carried out, the calibrated weight will be displayed with the
indication of the current unit of measurement.