Error Codes - Adam Equipment Solis SAB Série Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

13.0 ERROR CODES

ERROR MESSAGE
ON DISPLAY
ERR01
ERR02
ERR03
ERR04
ERR05
ERR06
ERR07
ERR08
ERR09
ERR10
ERR F
"UNLOAD"
"CAL But":
©Adam Equipment Co. Ltd. 2018
MEANING
Weight not stable after operation
of tare
impossible
to
calibration due to instability of
the balance
calibration weight not correct or
balance unstable
weight of samples for the part
counting function not adequate
or unstable
impossible to print because of
weight unstable
Weight cannot get stable in
density mode
Weight cannot get stable in
percentage weighing mode
Anomaly
on
auto-calibration
motor
Weight cannot get stable in
formulation mode
Weight of component out of
tolerance in formulation mode
Flash memory damaged
weight loaded on the pan or pan
not positioned properly
the
balance
recalibration
Over-range condition
Under-range condition
POSSIBLE SOLUTIONS
Protect the balance from air flows
or from vibrations from unstable
work surface.
start
the
Protect the balance from air flows
or from vibrations from unstable
work surface.
Calibrate with correct weight or
protect
environment
models with internal calibration
remove the screw in the left lower
part of the balance (see par5)
Select a bigger number of samples
or
protect
vibrations.
Protect
environment disturbance.
Protect
environment disturbance.
Protect
environment disturbance.
Contact service staff
Protect
environment disturbance.
Reduce quantity
Ask assistance from authorized
service staff
Remove the weight from the pan
or position properly the pan and
underpan.
requires
Unload weights, if any, on the pan,
and press the CAL button
Unload the weights loaded on the
pan
Locate the pan and underpan
correctly.
the
balance
from
disturbance.
the
balance
from
the
balance
from
the
balance
from
the
balance
from
the
balance
from
In
67

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Solis spb série

Table des Matières