2. When the value of the calibration weight begins to flash, load the calibration weight on the
weighing pan.
3. Wait for the calibrated weight to be displayed and the stability symbol to turn on, and then
remove the weight from the pan.
4. When "0.000" is shown on the display, press the PRINT button for a prolonged manner
until the beeping stops. The acquisition and automatic storage of the internal weight will
now begin. During the acquisition cycle, the display will show the following flashing text:
5. Once the internal calibration has been stored, the balance will return to the normal weighing
condition.
6. Now re-enter the calibration menu as described in paragraph 10.12.1 and set the desired
internal, automatic, or external calibration mode.