7. Press the tare, if necessary, and load the solid. Wait for the stability symbol to appear and
press the PRINT button to acquire the value. The word 'wait...' will appear while the weight
is being taken.
8. The weight of the solid in the liquid will then be requested. Carry out the tare of the drum
in the liquid. Put the solid in the drum, immerse the solid, and wait for the stability indicator
to appear. Then press the PRINT button. The word ''wait...'' will be displayed while the
weight is being taken.
9. The result of the density calculation of the solid will now be displayed. If the balance is
equipped with a printer, it will be possible to print the density value by pressing the PRINT
button. ("ERRO6" will be shown on the display in case of error)
10. Now press the ON/OFF button to exit from the density function, or the MENU button to
carry out the density measurement for another solid.