4.0 CALIBRATION 
 
 Press the [Zero] key and power up the scale using the on/off button. 
 Release the [Zero] key and the display will show "SPAN". Make sure the 
hook is empty unless you intend to use any chains, slings etc to hold the 
calibration  mass.  Add  any  items  such  as  these  to  the  hook  before 
continuing. 
 Press  the  [Zero]  button  and  the  display  will  show  "SET"  then  "00000" 
when the reading at zero is stable. This completes the zero calibration 
 Following zero calibration the value of the mass being used to calibrate 
the  scale  should  be  entered.  Use  the  [Hold]  button  to  increment  the 
right  hand  digit,  and  the  [Unit]  button  to  move  this  digit  to  the  left. 
Proceed  by  entering  a  number  for  the  next  digit  if  required,  until  the 
value of your chosen calibration weight is displayed correctly.  
 Load  this  mass  onto  the  hook  and  when  the  scale  is  stable  press  the 
[Zero] key. The display shows 'HOLD'. 
 After a few seconds, the calibration is complete and the scale returns to 
normal weighing showing the value of the calibration mass as used. 
 
**  To  calibrate  in  Lbs  make  sure  that  the  scale  is  Lbs  mode  before 
commencing,  to  calibrate  in  Kg  make  sure  the  scale  is  in  Kg  mode  before 
commencing. 
 