7.0 CALIBRATION

• Calibration is recommended for best accuracy. It is suggested that the
scale be calibrated when first installed and any time the scale is moved
to a different environment.
• To ensure best accuracy calibrate before use.
• Ensure weights are ready before calibrating, the weight should be of a
known accuracy, for example OIML class M1 or Class 4 of ASTM E617.
• Do not calibrate the scale if no calibration weight is available.
• Calibration can only be accessed if the Calibration function has been
enabled as described in section 6.0.
7.1

AUTOMATIC EXTERNAL CALIBRATION

• In the weighing mode, press [Func] key to access calibration mode. The
display will show "CAL".
• Press [Zero] key, an access code (Pin code) is required to continue, The
display will show "Pin" .
• To enter the access code press the [Func] key, [Unit] key and[Tare] key
in sequence, then press the [Zero] key to confirm the pin code.
• The display will show the weighing unit last used to calibrate the scale,
"S-Unit kg"(or "S-Unit Lb") ,press [Unit] key to change to lb's or kg's for
the calibration weighing unit value.
• Press [Zero] key to start the zero calibration. The display will show "CAL
0" and when the stable symbol is displayed press [Zero] key to confirm.
• The display will now show the calibration mass expected. "CAL XX Kg "
or "CAL XX Lb " .Press [Tare] key to change the calibration values if
necessary. Each model has 3 different kg and lb calibration selection
values.