Masimo RD SET DCI Série Mode D'emploi page 4

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 6
B. Attaching the sensor to the patient
1. Refer to Fig. 1a. Open the sensor by pressing on hinge tabs. Place the selected digit over the sensor window of the RD SET DCI or DCI-P. The fleshiest part of
the digit should be covering the detector window in the lower half of the sensor. The top half of the sensor is identified by the cable. On a finger site, the tip of
the finger should touch the raised digit stop inside the sensor. If the fingernail is long, it may extend over and pass the finger stop.
2. Refer to Fig. 1b. The hinged tabs of the sensor should open to evenly distribute the grip of the sensor along the length of the finger. Check position of sensor
to verify correct positioning. Complete coverage of the detector window is needed to ensure accurate data.
3. Refer to Fig. 1c. Orient the sensor so that the cable will be running towards the top of the patient's hand.
NOTE: With smaller digits, in order to completely cover the detector window, it may be necessary to use a lower weight range sensor. The sensor is not intended for use on the
thumb or across a child's hand or foot.
C. Attaching the Sensor to the Patient Cable
1. Refer to Fig. 2a. Properly orient the sensor connector and insert the sensor connector completely into the patient cable connector.
2. Refer to Fig. 2b. Close the protective latch completely.
D. Disconnecting the Sensor from the Patient Cable
1. Refer to Fig. 3a. Lift up the protective latch.
2. Refer to Fig. 3b. Pull firmly on the sensor connector to remove it from the patient cable.
NOTE: To avoid damage, pull on the sensor connector, not the cable.
CLEANING
Remove the sensor from the patient and disconnect it from the patient cable.
1.
Clean the sensor by wiping with a 70% isopropyl alcohol pad.
2.
Allow the sensor to dry thoroughly prior to placement on a patient.
3.
or
If low-level disinfection is required, use a 1:10 bleach / water solution.
1.
Saturate a cloth or gauze pad with the cleaning solution and wipe all surfaces of the sensor and cable.
2.
Saturate another cloth or gauze pad with sterile or distilled water and wipe all surfaces of the sensor and cable.
3.
Dry the sensor and cable with a clean cloth or dry gauze pad.
4.
CAUTIONS:
Do not use undiluted bleach (5% - 5.25% sodium hypochlorite) or any cleaning solution other than those recommended here because permanent damage to the sensor
may occur.
Do not immerse the sensor or connector in any liquid solution.
Do not sterilize by irradiation, steam, autoclave or ethylene oxide.
SPECIFICATIONS
When used with Masimo SET pulse oximetry monitors or with licensed Masimo SET pulse oximetry modules RD SET DCI® and DCI-P reusable sensors have the following
specifications:
Used with:
Sensor
Body Weight
+70 C
-40 C
+1060 hPa - +500 hPa
795 mmHg - 375 mmHg
Application site
NON
STERILE
SpO
Accuracy, No Motion, (70 - 100%
2
5%-95% RH
SpO
Accuracy, Motion
2
LATEX
SpO
Accuracy, Low Perfusion
2
95%
4
%
Pulse Rate
Accuracy, No Motion, (25 - 240 bpm)
5%
Pulse Rate Accuracy, Motion
Title: Graphics, Sensor/Cable symbols, 03/06
Pulse Rate Accuracy, Low Perfusion
GR-14231
DRO-13914
NOTE: The A
Accuracy is calculated based upon measurement values that are statistically distributed; approximately 68% of the measured values fell within +/- the Arms
rms
value when compared to the reference device under a controlled study.
1
The Masimo SET Technology has been validated for no motion accuracy in human blood studies on healthy adult male and female volunteers with light to dark pigmented skin in
induced hypoxia studies in the range of 70%-100% SpO
population.
2
The Masimo SET Technology has been validated for motion accuracy in human blood studies on healthy adult male and female volunteers with light to dark pigmented skin in induced
hypoxia studies while performing rubbing and tapping motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in
induced hypoxia studies in the range of 70%-100% SpO
population
3
The Masimo SET Technology has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and Masimo's simulator with signal strengths of
greater than 0.02% and transmission of greater than 5% for saturations ranging from 70% to 100%. This variation equals plus or minus one standard deviation which encompasses 68%
of the population.
4
The Masimo SET Technology has been validated for pulse rate accuracy for the range of 25-240 bpm in bench top testing against a Biotek Index 2 simulator and Masimo's simulator
with signal strengths of greater than 0.02% and transmission of greater than 5% for saturations ranging from 70% to 100%. This variation equals plus or minus one standard deviation
which encompasses 68% of the population.
COMPATIBILITY
This sensor is intended for use only with devices containing Masimo SET oximetry or pulse oximetry monitors licensed to use RD SET sensors. Each sensor is
designed to operate correctly only on the pulse oximetry systems from the original device manufacturer. Use of this sensor with other devices may result in no
or improper performance.
For Compatibility Information Reference: www.Masimo.com
WARRANTY
Masimo warrants to the initial buyer only that these products, when used in accordance with the directions provided with the Products by Masimo, will be free of defects in
materials and workmanship for a period of six (6) months. Single use products are warranted for single patient use only.
THE FOREGOING IS THE SOLE AND EXCLUSIVE WARRANTY APPLICABLE TO THE PRODUCTS SOLD BY MASIMO TO BUYER. MASIMO EXPRESSLY DISCLAIMS ALL OTHER ORAL,
EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION ANY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE. MASIMO'S SOLE
OBLIGATION AND BUYER'S EXCLUSIVE REMEDY FOR BREACH OF ANY WARRANTY SHALL BE, AT MASIMO'S OPTION, TO REPAIR OR REPLACE THE PRODUCT.
RD SET DCI
> 30 kg
Finger
1
)
2
3
3 bpm
4
5 bpm
4
3 bpm
against a laboratory co-oximeter. This variation equals plus or minus one standard deviation which encompasses 68% of the
2
against a laboratory co-oximeter. This variation equals plus or minus one standard deviation which encompasses 68% of the
2
Masimo SET Technology
RD SET DCI-P
Finger or toe
2%
3%
2%
4
10 - 50 kg
2%
3%
2%
3 bpm
5 bpm
3 bpm
8994B-eIFU-0318

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Rd set dci-p série

Table des Matières