Instructions for Use
Connection of the High Vacuum Flange

Installation of ISO-K Flange

For ISO-K flange connections, fix the two flanges with the clamps model
IC63250DCMZ or equivalent, as shown in the following figure (a protective
screen can optionally be used).
Figure 3
Use the required number of claw clamps:
Tighten the claw clamps with a torque of 22 Nm.
202/258
4 for ISO 100 flange
4 for ISO 160 flange