6 Verification and Calibration
Verification Setups
Test Considerations
For optimum performance, all verification and calibration procedures should comply with the
following:
Ambient temperature is stable, within 23 °C ±2°C.
l
Ambient relative humidity is less than 80%.
l
30 minute warm-up period before verification or adjustment.
l
Cables as short as possible, twisted or shielded to reduce noise.
l
282
Keysight RP7970 & RP7980 Series Operating and Service Guide