CALibrate:DATE <"date">
CALibrate:DATE?
Enters the calibration date in nonvolatile memory. Enter any ASCII string up to 15 characters. The
query returns the date.
Parameter
<"date">
String program data. Enclose string parameters in single
or double quotes.
Enters the calibration date: CAL:DATE "12/12/12"
CALibrate:LEVel P1|P2|P3
CALibrate:LEVel?
Advances to the next level in the calibration. P1 is the first level; P2 is the second; P3 is the third.
Parameter
P1|P2|P3
Selects the first calibration point: CAL:LEV P1
Some calibration sequences may require some settling time after sending CAL:LEV but before read-
l
ing the data from the DVM and sending CAL:DATA.
CALibrate:PASSword <password>
Sets a numeric password to prevent unauthorized calibration. This is the same as the
Parameter
<password>
A numeric value of up to 15 digits
Set a new password to a value of 1234: CAL:PASS 1234
If the password is set to 0, password protection is removed and the ability to enter calibration
l
mode is unrestricted. The as-shipped setting is 0 (zero).
To change the password: unsecure calibration memory with old code, then set the new code.
l
If you forget your password, refer to
l
This setting is non-volatile; it will not be changed by power cycling or *RST.
l
Keysight RP7970 & RP7980 Series Operating and Service Guide
Typical Return
<last cal date>
Typical Return
(none)
Typical Return
(none)
Calibration
Switches.
5 SCPI Programming Reference
Admin
password.
193