4 Using the Regenerative Power System
Front Panel Menu Reference
Select Transient\Control.
Select Trigger to generate an immediate trigger signal
regardless of the trigger source setting.
If a digital pin is configured as the trigger source, the instrument will wait indefinitely for the trigger
signal. If the trigger does not occur, you must manually return the trigger system to the idle state. The
following commands return the trigger system to the idle state:
Front Panel Menu Reference
Select Transient\Control.
Then select the Abort control.
When a trigger is received, the triggered functions are set to their programmed transient values. When
the triggered actions are completed, the trigger system returns to the idle state.
You can test the TRAN-active bit in the Operation Status register to know when the transient trigger
system has returned to the idle state.
Front Panel Menu Reference
Select Transient\Control.
The Trig state field indicates "Idle".
If a bit value of 64 is returned in the query, the TRAN-active bit is true, and the transient action is NOT
complete. When the TRAN-active bit is false, the transient action is complete. Refer to
for more information.
132
SCPI Command
To generate a transient trigger:
TRIG:TRAN
Alternatively, if the trigger source is BUS, you can
also program a
*TRG
or an IEEE-488 <get>
command.
SCPI Command
ABOR:TRAN
SCPI Command
To query the TRAN-active bit (bit 6):
STAT:OPER:COND?
Keysight RP7970 & RP7980 Series Operating and Service Guide