Télécharger Imprimer la page

Keysight RP7971A Guide D'utilisation Et De Maintenance page 240

Système d'alimentations régénératrices

Publicité

5 SCPI Programming Reference
STATus:PRESet
Presets all Enable, PTR, and NTR registers.
Operation register
STAT:OPER:ENAB
STAT:QUES<1|2>:ENAB
STAT:OPER:NTR
STAT:QUES<1|2>:NTR
STAT:OPER:PTR
STAT:QUES<1|2>:PTR
Parameter
(none)
Preset the Operation and Questionable registers: STAT:PRES
STATus:QUEStionable<1|2>[:EVENt]?
Queries the
event register
stores (latches) all events that are passed by the Operation NTR and/or PTR filter. Reading the
Questionable Status Event register clears it.
Parameter
(none)
Read questionable status event register #1: STAT:QUES1?
The value returned is the binary-weighted sum of all enabled bits in the register.  For example, with
l
bit 2 (value 4) and bit 4 (value 16) set, the query returns +20.
*RST has no effect on this register.
l
STATus:QUEStionable<1|2>:CONDition?
Queries the
condition register
holds the live (unlatched) operational status of the instrument. Reading the Questionable Status
Condition register does not clear it.
Parameter
(none)
Read questionable status condition register #1: STAT:QUES1:COND?
The value returned is the binary-weighted sum of all enabled bits in the register.  For example, with
l
bit 2 (value 4) and bit 4 (value 16) set, the query returns +20.
The condition register bits reflect the current condition. If a condition goes away, the cor-
l
responding bit is cleared.
*RST clears this register, other than those bits where the condition still exists after *RST.
l
240
Questionable register
all defined bits are disabled
all defined bits are disabled
all defined bits are enabled
for the
Questionable Status
for the
Questionable Status
Keysight RP7970 & RP7980 Series Operating and Service Guide
Preset setting
Typical Return
(none)
group. This is a read-only register, which
Typical Return
<bit value>
group. This is a read-only register, which
Typical Return
<bit value>

Publicité

loading