4 Using the Regenerative Power System
Enable the output transient function
First, you must enable the output to respond to transient triggers. Unless an output transient function
is enabled, nothing will happen even if you have programmed the transient parameters and generated
a transient trigger.
Front Panel Menu Reference
Select Transient\Mode.
If you are operating in voltage priority mode select
Voltage mode. If you are operating in current priority
mode select Current mode. If you are operating in
power priority mode, select Power mode.
In the dropdown list, select Step, List or Arb transients.
Then press Select.
In Step mode, the triggered value becomes the immediate value when the trigger is
received. In Fixed mode, trigger signals are ignored; the immediate values remain in
effect when a trigger is received.
Program the transient parameters
For example, set the triggered voltage level if you are programming a voltage step:
Front Panel Menu Reference
Select Transient\Step.
Select the Trig Voltage box to set the voltage.  Enter a
value and press Select.
Select the trigger source
A TRIGger:TRANsient[:IMMediate] command over the bus will always generate an
immediate transient trigger, regardless of the selected trigger source.
Unless you are using the front panel menu or a TRIGger:TRANsient[:IMMediate] command to trigger
the transient, select a trigger source from the following:
Trigger Source
Bus
Selects GPIB device trigger, *TRG, or <GET> (Group Execute Trigger).
External
Selects ANY pin that has been configured as a trigger input on the digital control port.
Immediate
Triggers the transient as soon as it is INITiated.
Pin<1-7>
Selects a specific pin<n> that is configured as a Trigger Input on the digital control port.
130
SCPI Command
To enable the transient function, use:
VOLT:MODE STEP
VOLT:MODE LIST
VOLT:MODE ARB
or
CURR:MODE STEP
CURR:MODE LIST
CURR:MODE ARB
or
POW:MODE STEP
POW:MODE LIST
POW:MODE ARB
SCPI Command
To set a voltage step level of 15 V use:
VOLT:TRIG 15
Description
Keysight RP7970 & RP7980 Series Operating and Service Guide