Hemostat Setup Screen Description:
² "Pressure": when you use hemostat function, you need to preset a cuff pressure for hemostasis. The pressure is
adjustable, and its adjusting limit is different for different patient category:
for neonates: preset range: 70~100 mmHg, default value: "90" mmHg;
for pediatrics: preset range: 80~130 mmHg, default value: "110" mmHg;
for adults: preset range: 80~180mmHg, default value: "140" mmHg.
If the pressure drops down slowly under 10mmHg compared with the preset value due to little air leakage in
the pneumatic system when time passes by, the monitor will re-inflate to maintain the cuff pressure close to the
preset pressure value.
Note: the unit of cuff pressure is the same with the NIBP unit in NIBP Setup.
² "Duration": After presetting the cuff pressure, you need to set the time period for maintaining the preset
pressure after inflation. "5, 6, 7,...120" minutes adjustable. The default value is "40" minutes.
If the set value is "xx" minutes, the monitor will count down from "xx" minutes automatically when starting
cuff inflation. When time is up, it will deflate automatically.
² "Alert Time": the alert time for reminding user that the operation of hemostat is going to be end after this time
period. 1 to 60 minutes adjusting range with 1 minute step, the default value is "5" minutes. If the set value is
"xx" minutes, the monitor will produce alarm sound until ending deflation when counting down time reaches to
"xx" minutes. The alarm type is high priority alarm. (For example: the duration is 40 minutes, the alert time is 5
minutes, the alarm will ring for prompt when the duration counting down to 5 minutes. The Prompt Info area
starts to prompt: Hemo C-D 300 seconds. )
"Start": shift cursor to "Start" and press "
being inflated; Pressing "Stop" button can stop using this function. After deflation, it will change to "Start" again.
4.9.6 Patient Info
Screen description:
²
"Patient ID": change or set current patient's ID number, 0~100 adjustable; Once the patient ID is changed, the
history data in trend graph will be cleared, and the parameter settings will be resumed to default value.
²
"Category": change or set the category of current patient; three options "adult", "pediatric" and "neonate",
press OK key to confirm the setting and the patient type indicator will be ON at the front panel of the device.
The default is "adult".
" key, "Start" becomes "Stop" and meanwhile the blood cuff starts
Figure 4.18 Patient Info. setup screen
35