Guidance and manufacturer's declaration-electromagnetic immunity
Gima Vital Pro Monitor is intended for use in the electromagnetic environment specified below. The customer or
the user of the equipment or system should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge(ESD)
IEC61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC61000-4-11
Power frequency
(50Hz/60Hz) magnetic
field
IEC61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
for all EQUIPMENT AND SYSTEMS
IEC60601 test level
±8 kV contact
±15kV air
±2kV for power
Supply lines
±1 kV for
input/output lines
±1kV line (s) to line(s)
±2kV line(s) to earth
3A/m
Table 2
Electromagnetic
Compliance level
-guidance
±8 kV contact
Floors should be wood, concrete
±15kV air
or ceramic tile. if floors are
covered with synthetic material,
the relative humidity should be at
least 30%
±2kV for power
Mains power quality should be
Supply lines
that of a typical commercial or
±1 kV for
hospital environment.
input/output lines
±1kV differential
Mains power quality should be
mode
that of a typical commercial or
±2kV common
hospital environment.
mode
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the equipment or system
requires continued operation
during power mains interruptions,
it is recommended that the
equipment or system be powered
from an uninterruptible power
supply or a battery.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
3A/m
commercial or hospital
environment.
51
environment