9.1 No Display on the Screen
Shut down the monitor and unplug the power. Use a universal meter to check whether the outlet has proper voltage, if the
power cable is in good condition, and that the power cable is properly connected with this apparatus or outlet. Remove
the fuse from the back cover of the monitor, and make sure it is in good condition.
9.2 Excessive ECG Signal Interference or Too Thick Baseline
1. Check if the plate electrodes are properly located, and if valid plate electrodes are used.
2. Check whether the lead wires are inserted properly. If no ECG curve is displayed, check if the ECG lead wires
are broken.
3. Make sure the mains outlet has standard grounding wire.
4. Check if the grounding wire of the apparatus is properly grounded.
9.3 No Blood Pressure and Pulse Oxygen Readings
1. Check that the blood pressure cuff is properly wrapped around the arm according to the operating instructions,
that the cuff does not have a leak, that connections are secure on the cuff and tubing and that the inlet is closely
connected with the NIBP jack on the side panel. Check that the LED of SpO
oxygen probe is properly connected to the SpO
2. If the problems still exist, please contact the local dealer.
9.4 Blank Print-out
1. Check whether the printing paper is installed the right way around (i.e. with the sensitive side upwards). Please
reinstall itif necessary.
2. If the problems still exist, please contact the local dealer.
9.5 System Alarm
1. When the parameter value is higher or lower than the alarm limits, the alarm will sound. Please examine the
condition of the patient and check whether the alarm limit values are set appropriately.
2. Probe off. Please check the connection of the probes.
Note: If a problem arises with this machine in service, follow the instructions below to attempt to eliminate the
problem first. If the attempt fails, contact the dealer in your local area or the manufacturer. Do not open the
monitor casing without permission.
Chapter 9 Troubleshooting