vascular contracting drugs, the SpO
more sensitive to interference.
²
For those with a substantial amount of staining dilution agent (such as methylene blue, indigo green and acid
indigo blue), or carbon monoxide hemoglobin (COHb), or methionine (Me+Hb) or thiosalicylic hemoglobin,
and some with icterus problem, the SpO
²
Drugs such as dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor resulting in
serious SpO
2
²
As the SpO
2
measurement result of some patients with serious anemia may also present as good SpO
8.3.5 Points to be noted in SpO
²
The finger should be properly placed (see the illustration to follow in this instruction manual), or else it may
cause inaccurate measurement results.
²
Make sure that the sensor is lined up so the red and infrared LEDs pass through the capillary arterial vessels .
²
The SpO
sensor should not be used at a location or limb with an arterial or blood pressure cuff attached or
2
receiving intravenous injection.
²
Do not fix the SpO
inaccurate measurement result of SpO
²
Make sure the optical path is free from any obstacles such as adhesive tape.
²
Excessive ambient light (such as fluorescent lights, infrared heaters and direct sunlight) may affect the
measuring result.
²
Strenuous activity of the patient or extreme electrosurgical interference may also affect the accuracy.
²
Please do not use the SpO
²
Always observe the plethysmogram (waveform), which is auto-scaled within the range of 100. If the
waveform is not smooth or is irregular, this may indicate that the SpO
rely on your clinical judgment, rather than the monitor readout.
²
A functional tester cannot be used to assess the accuracy of the pulse oximeter monitor or a SpO
However, a functional tester, such as SpO
oximeter is reproducing the given calibration curve. Before testing the oximeter, please determine that the
appropriate calibration curve is being used. If necessary, request this from the manufacturer and download it
into the test device.
8.4 Temperature Monitoring (optional)
The sensor is thermo-resistor type (25℃ 5kΩ) with constant micro current. Calculating the temperature measurement
from the voltage. Temperature measurement can be obtained via two method: measure through body surface temperature
and through the temperature inside the body cavity (with oral or rectal placement of the probe).
Normal value: body surface: 36.5℃~37℃; inside body cavity: 36.5℃~37.7℃
Notes:
Ø
Attach the TEMP transducer/sensor to the patient; ensuring good contact with the skin if this method is chosen.
Secure the sensor with adhesive tape.
Ø
Especially for pediatric patient, they like sports, pay more attention to the transducer fixing.
waveform (PLETH) will decrease. In this case, the measurement will be
2
measurement errors.
value serves as a reference value for judgment of anemic anoxia and toxic anoxia, the
and Pulse Measuring
2
sensor with adhesive tape, which may result in venous pulsation and consequential
2
sensor when having the MRI, or burn may be caused by faradism.
2
readings may be inaccurate.
2
.
2
simulator can be used to check how accurately a particular pulse
2
59
value.
2
readings are not accurate. If in doubt,
2
sensor.
2