(such as SpO
, NIBP...) to make the high/low limit setting.
2
☞
Limits setup: Move to the High or Low limits of the alarm settings, and press the "Alarm" silence key to turn ON
or OFF the alarm for the setting. The alarm silence indicator will reflect this with a yellow light showing that
alarms are silenced.
Refer to Chapter 11.2 for detailed default alarming values of all parameters and setup range.
M Whenever the monitor is used, check the alarm limits to ensure that they are appropriate
for the patient being monitored.
M When mains power supply is lost for not more than 30s, the alarm settings prior to the
power interrupt will sustain or be restored automatically.
5.5 Verifying Alarm Function
To verify the effectiveness of the alarm function, set the monitor working at "Demo" mode in system parameter settings
menu. Adjust the alarm limits or change alarm setting, then pay a close attention to the alarm signal. If the visual and
audible alarm indications appear according to your setting, it means the alarm function is effective. Do NOT set the
alarm volume lower than the background noise.