Safety Instructions
The Electrosurgical Generator contains no user serviceable parts. To avoid risk of
electric shock and equipment malfunction do not remove any cover.
Modification or opening of the generator by unqualified persons may result in
electrical shock and will breach warranty of the generator.
Microwave energy should be not be applied to persons wearing metallic jewelry or
clothing containing metallic material. Hearing aids should be removed. Patients with
implanted electronic devices and/or electrodes should be excluded from treatment
with microwaves and from areas where the microwave equipment is operated. For
patients with cardiac pacemakers or other active implants, a possible hazard exists
due to interference with the action of the pacemaker or the pacemaker may be
damaged, and when in doubt, approved qualified advice should be obtained.
Caution
Disconnecting the Instrument from the Interface Cable while cut or coagulation are
activated poses the risk of burns when touching the connector of the Interface
Cable. Avoid touching the connectors of the Interface Cable at all times.