Normal Use
Adjusting Power Outputs – Using the Footswitch Pedals
TYPE 1 & 2 Instruments Only
Press and hold the footswitch Menu Button (the round button in the center of
STEP 1
the footswitch) until the speaker of the Electrosurgical Generator sounds an
audible acknowledgement; then release the button that was applied.
The Electrosurgical Generator Display will
then highlight the cut and/or coagulation
STEP 2
power settings according to the selected
instrument. The display for RS2 is
illustrated as an example.
Pressing and releasing the yellow (Left) footswitch pedal repeatedly will cycle
the Display through the range of RF cut power settings that are available for the
connected Instrument.
STEP 3
Pressing and releasing the blue (Right) footswitch pedal repeatedly will cycle the
Display through the range of coagulation power settings that are available for
the connected Instrument.
Once the power output settings have been selected, press and hold the center
STEP 4
footswitch Menu button until the Electrosurgical Generator speaker sounds an
audible acknowledgement.
Releasing the footswitch Menu button then returns the Display of the
Electrosurgical Generator to the new Treatment settings selected.
STEP 5
Before commencing treatment always confirm that the correct power settings
have been made and are displayed accordingly.
RS2
CUT 30
CUT 30
COAG 09
COAG 09
Page 17 of 248