Output, RF (Treatment
Mode):
Output, Microwave
(Treatment Mode):
Physical:
Standards Compliance:
Output Frequency:
Power Capability:
Rated Load:
Recommended Duty Cycle:
RF Crest Factor:
Maximum Voltage:
Power Capability:
Output Frequency:
Rated Load:
Recommended Duty Cycle:
(Please refer to applicable
IFU):
(The duration of the "ON"
period is automatically
limited by the generator and
specific accessory, re-
application of the blue
footswitch pedal is required
to resume energy activation)
RF Crest Factor:
Software Version:
Weight:
Dimensions:
Basic Safety and Essential
Performance:
Collateral Standard, EMC:
High Frequency Surgical
Devices:
Microwave Therapy
Equipment:
Endoscopic Equipment:
396.7 kHz +/-2.0 kHz
35 W (Max)
400 Ohm
10 s ON, 30 s OFF, 1 hour
1.6 continuous to 3.7 pulsed
460 V peak
62 W nominal at maximum setting measured at the
output socket
5,800 MHz +/- 1 MHz
50 Ohm
RS2 - 10 s ON, 1 s OFF x 3 then 268 s OFF (1-hour
max operation time)
HS1 – 10 s ON, 1 s OFF x 5 then 246 s OFF (1-hour
max operation time)
RG1 -10 s ON, 1 s OFF x 3 then 268 s OFF (1-hour
max operation time)
RGS1 -10 s ON, 1 s OFF x 3 then 268's OFF (1-hour
max operation time)
AB1 – 180 s ON, 30 s OFF x 9 then 28 min 30 sec
OFF (1-hour max operation time).
NP1 – 120 s ON, 120 s OFF x 8 then 28 min OFF (1-
hour max operation time).
1.4 (output is a sine wave)
The software version is displayed on the
Electrosurgical Generator following switch on from
Standby
16 kg / 35 lb
Height 165 mm +/- 5 mm
Width 400 mm +/- 5 mm
Depth 458 mm +/- 5 mm
IEC 60601-12:2005 + A1
IEC 60601-1-2:2014
IEC 60601-2-2:2017
IEC 60601-2-6:2012
IEC 60601-2-18:2009
Page 30 of 248