Normal Use
Connect the required Instrument to the Interface cable.
Select and Confirm Connected Instrument
Warning
Ensure the correct Instrument is chosen from the options displayed by the
generator. The wrong choice may result in serious injury to the patient and damage
to the connected instrument.
When an instrument has been
connected the generator will
display the Select Instrument
screen as illustrated.
STEP 1
When the Select Instrument screen is first displayed, '===' is highlighted
indicating that no instrument is selected. Use the '+' or '-' control button (4)
to highlight the connected instrument then press the menu control button
(5) or footswitch menu button (the round button in the center of the
footswitch) to select the connected instrument.
The generator will then display
the Confirm Instrument screen.
This is illustrated below when RS2
was selected from the Select
Instrument screen.
Use the '+' or '-' control button (4) to highlight the instrument, then press
STEP 2
the menu control button (5) or footswitch menu button to confirm that the
correct instrument has been selected.
If the wrong instrument has been selected, use the '+' to highlight '===' then
press the menu control button (5) or footswitch menu button to return to
the Select Instrument screen.
If the wrong instrument has been selected and confirmed so that the
Treatment Screen is displayed, disconnect the instrument and then
reconnect the instrument. The Select Instrument screen will then be
displayed.
Scroll to Select Instrument
===
===
HS1
HS1
AB1
RS2
NP1
RG1
RGS1
Scroll to Confirm Instrument
===
===
RS2
Page 14 of 248