Cleaning
Caution
Cleaning does not render the Electrosurgical Generator surfaces sterile.
Disconnect the Electrosurgical Generator from the AC supply before cleaning to
avoid risk of electrical shock.
Completely remove and allow all cleaning liquids to dry before connecting the
Electrosurgical Generator to the AC supply to avoid risk of electrical shock.
The Electrosurgical Generator, AC power cord and footswitch should be cleaned before and
after each use in accordance with procedures or other requirements of the hospital or other
clinical site.
The external surfaces should be cleaned only by wiping with a clean, non-abrasive disinfecting
wipe that is lightly dampened with the following:
70/30 Isopropyl Alcohol (IPA)/Water mix.
In the event of any damage or degradation occurring to the Electrosurgical Generator as a
result of cleaning, please contact Creo Medical immediately for advice. Do not use the
Electrosurgical Generator when it appears to be damaged.
Refer to the manufacturer's instructions for cleaning of the Footswitch.