Trouble Shooting
Message:
"Release
Footswitch"
Message:
"Attach
Instrument
When Ready to
Proceed"
Message:
"Incorrect Pedal
Applied"
Audible output is
too loud or to
quiet
Display is too
bright or too dim
Message: "No
cut detected,
release pedal
and reapply"
For Instruments
intended for
endoscopic use,
only:
During cutting
there are image
artefacts on the
endoscopic
monitor
One of the pedals is pressed during the self-tests following turning ON
the Electrosurgical Generator – release the pedal.
Actions to be taken are:
• If no Instrument is connected, connect the required Instrument.
• If no Interface Cable is connected, connect the Interface Cable
and then connect the required Instrument to the Interface Cable.
• If an Interface Cable and an Instrument are both connected,
check all connections.
• If the error condition persists, replace the Instrument. Return the
Instrument to Creo Medical.
• If the error condition persists, replace the Interface Cable.
• If replacing the Instrument and the Interface Cable does not
correct the error condition, then the Electrosurgical Generator is
most likely to be faulty. Contact Creo Medical.
Applicable only when using a single energy type instrument, for
example, HS1 that uses the Coag output only.
Message will be displayed when the wrong foot pedal has been
pressed – release the wrong pedal and press the correct pedal.
Adjust the Speaker Volume as described in "Additional Settings".
Adjust the Screen Brightness as described in "Additional Settings".
Release the Footswitch pedal and then reapply. If the message is still
showing, remove the Instrument and clean the tip of the Instrument.
Refer to the instruction for use of the Instrument for cleaning
instructions.
If this fault condition occurs repeatedly, replace Instrument.
There might be electrical interference between the Electrosurgical
Generator and the endoscopic visualization equipment, or light
generated from around the Instrument during cutting could be causing
saturation of the video camera. Furthermore, excessive tissue build-up
on the tip of the Instrument increases the likelihood of arcing and
bright flashes that temporarily degrade video quality.
Release the footswitch and then reapply and / or reposition
endoscopic equipment connection cables away from the
Electrosurgical Generator and its Instrument cables or reposition the
Electrosurgical Generator and clean the tip of the Instrument.