Normal Use
Connecting the Instrument
Warning
Do not use the Interface Cable beyond its lifetime as specified in the Instruction for
Use of the Interface Cable.
The performance of the Interface Cable degrades if used after its specified lifetime
which may result in unintended tissue effect.
Ensure that the Interface Cable is securely connected to the Electrosurgical
Generator and the Instrument before activating the output of the Electrosurgical
Generator.
Incorrect connection poses the risk of the user being exposed to high voltages or
degradation of delivered power and unintended tissue effects.
The Electrosurgical Generator, Instruments, footswitch, and their connection cables
should be positioned such that damage or accidental activation is avoided and that
they do not present a tripping hazard.
The Interface Cable is required to connect an Instrument for use.
The Interface Cable is connected to the Electrosurgical Generator Output connection socket (9)
located on the Front Panel.
Refer to the Instructions for Use provided with the Interface Cable before
connecting it to either the Electrosurgical Generator or to the Instrument
Refer to Instructions of Use for the Instrument before connecting it to the Interface
Cable.