Normal Use

Warning
Do not activate the output when the Instrument is in contact with or in close
proximity to metallic objects or other medical devices.
Do not use in combination with other electrosurgical generators.
Contact with other medical devices may cause burns to the patient.
Avoid touching the Output connection socket (9), connectors of the Interface Cable,
and the tip of the Instrument at all times.
The RF output sends a RF signal once per second to detect the connection of an
instrument that may be noticeable when touching the Output connection socket.
Energy for cutting and coagulation should be delivered at the minimum power and
for the minimum time necessary to complete the procedure. Use of excessive energy
and/or application for extended time, may result in patient injury due to excessive
penetration and lateral spread of thermal effects.
Turning ON the Electrosurgical Generator
To turn ON the Electrosurgical Generator, on the Rear Panel push the part
of the On / Off Power switch (10) marked 'I' inwards.
Immediately after turning ON the Electrosurgical Generator, the red
warning indicator LED (2) on the Front Panel of the Electrosurgical
STEP 1.
Generator will illuminate, briefly turn off, and then illuminate again for a
few more seconds. Thereafter the red Warning indicator LED (2) will
switch off and the green On / Standby LED (1) will flash continuously
(note: that the Display will remain blank). The Electrosurgical Generator is
now in Standby Mode.
Press the ON / Standby button (1) on the Front Panel of the Electrosurgical
Generator to exit from Standby Mode. An audible tone will then be
generated from the internal speaker of the Electrosurgical Generator, the
STEP 2.
5 LEDs on the Front Panel will illuminate in turn and together, and the
alphabet will be displayed in 2 font sizes to confirm the Display screen is
working properly.
The Electrosurgical Generator continues to perform a series of internal
STEP 3.
self-checks while a screen is being displayed identifying Creo Medical
Electrosurgical Generator and the selected language.
When the self-checks are successfully completed, the message "Attach
STEP 4.
Instrument when ready to proceed" is displayed.
Page 12 of 248