Wipe disinfection for the camera and mirror sleeve

NOTE
Do not spray the CEREC Omnicam or immerse it in cleaning agents
or disinfectants!
1. Carry out a preliminary cleaning process (see "Pre-cleaning of the
mirror sleeve").
2. Use a new absorbent cotton gauze which is moistened with one of
the agents listed in the "Cleaning agents and disinfectants" section.
First wipe the camera enclosure and then the mirror sleeve. The
camera enclosure and mirror sleeve are disinfected through this
step.
3. Wipe down the mirror sleeve afterwards again with the absorbent
cotton gauze dipped in drinking water.
4. Next dry the mirror sleeve using a lint-free cloth.
8.1.5.8

High-level disinfection of the mirror sleeve (with HLD Set)

NOTE
Dry heat sterilization and high-level disinfection must not be
combined.
The complete process for high-level disinfection is as follows:
With HLD Set
A HLD set to support the HLD process should be ordered from Dentsply
Sirona with REF 66 05 120.
A
HLD container
B
Lid
1. Pre-cleaning (see „Pre-cleaning of the mirror sleeve  [→ 56]").
2. Put on Personal Protective Equipment at the latest when handling
the high-level disinfectant.
3. Remove the mirror sleeve from the CEREC Omnicam (see
„Removing the mirror sleeve  [→ 55]").
8.1 Care, cleaning, disinfection and sterilization
C
Metal holder
D
Plug
8 Maintenance
57