Technician Diagnostics - Dentsply Sirona CEREC AC Connect Notice D'utilisation

Masquer les pouces Voir aussi pour CEREC AC Connect:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Dentsply Sirona
Operating Instructions for the Acquisition Unit - With CEREC Omnicam
66 37 479 D3492
D3492.201.07.28.02    09.2019
Test result: Fail
An error was found on the PC-specific hardware. Replacing the PC
component or the PC may be required.
An image restore is not advisable.
1. Check to find out which components did not pass the test.
2. Write down the defective test number and, if available, the error
code and inform the technician of this as soon as possible.
3. Carry out the appropriate steps in the chapter entitled
Troubleshooting  [→ 84] .
End "Customer Diagnostics" test
1. Remove the PC diagnostic tool CD from the drive.
2. Switch the PC off by briefly pressing the on/off key.
10.3.2

Technician Diagnostics

ü You have started the "Technician Diagnostics" test and the
system configuration is displayed.
1. Using the information in the system configuration, check whether
the system corresponds to the default settings. You can scroll up
and down using the arrow keys.
Tip: The test starts automatically after 3 minutes if you do not press
an arrow key.
2. Press the "Esc" key on the keyboard.
10 PC Diagnostic Tool
10.3 Test procedure
81