Measuring the position of the exit window
1. Mount the bottom side of the calibration set to the tip of the camera.
2. Click on the "OK" button.
Ä The calibration process is continued.
Ä Once the calibration is complete, a message is displayed
indicating this.
3. Confirm the message by clicking the "OK" button.
Ä The CEREC Omnicam is calibrated.
Error message during calibration
The software indicates if an error occurs during calibration. If the
calibration process resulted in errors, restart the process.
End calibration
ü The software indicates that the calibration was completed
successfully.
➢ Click on the "OK" button.
Ä The CEREC Omnicam is calibrated.
8 Maintenance
8.2 Calibrating CEREC Omnicam