Disinfecting the camera body and mirror sleeve
8.1.5.7
NOTICE
Do not spray the CEREC Omnicam or immerse it in cleaning agents or
disinfectants!
➢ Use a new cotton gauze moistened with one of the cleaning products
listed in the „Cleaning and disinfection agents [ → 48]" section. Wipe
the camera body first and then the mirror sleeve. This step disinfects
the camera body and the mirror sleeve. Then dry the mirror sleeve
with a clean lint-free cloth.
8.1.5.8
High-level disinfection of the mirror sleeve (with HLD Set)
NOTICE
Dry heat sterilization and high-level disinfection must not be combined.
The complete process for high-level disinfection is as follows:
With HLD Set
A HLD set to support the HLD process should be ordered from Sirona with
REF 66 05 120.
A
HLD container
B
Lid
1. Pre-cleaning (see „Pre-cleaning of the mirror sleeve [ → 52]").
2. Put on Personal Protective Equipment at the latest when handling the
high-level disinfectant.
3. Remove the mirror sleeve from the CEREC Omnicam (see
„Removing the mirror sleeve [ → 51]").
C
Metal holder
D
Plug