8 Maintenance
8.1 Care, cleaning, disinfection and sterilization
64
Operating Instructions for the Acquisition Unit - With CEREC Omnicam
8.1.5.11

Use of CEREC Omnicam Disposable Sleeves

The CEREC Omnicam can use disposable plastic sleeves to ensure
maximum infection control. These are available with order number
66 32 264.
NOTE
The disposable sleeves cannot be used with the CEREC AF and
CEREC AI systems or when using the shade detection feature or
using the Ortho software.
1. Following each patient, remove the disposable sleeve from the
CEREC Omnicam camera and dispose of it according to the
standard procedure.
2. Wipe disinfection of the camera (see "Wipe disinfection for the
camera and mirror sleeve  [→ 57]").
3. Remove one disposable sleeve from its packaging. Slide the
disposable sleeve up to the stop on the mirror sleeve, so that it is
securely positioned on the CEREC Omnicam camera. Hold the
camera in other places, not just on the sleeve in order to prevent
the camera from falling.
4. Position the camera in the holder, so that it can warm up for 15
minutes prior to the intraoral exposure. Make sure that the holder
has been disinfected (see „Non-critical surfaces excluding monitor
screen  [→ 53]").
5. In the case of intraoral exposures with disposable sleeves, the
CEREC Omnicam camera should be positioned closer to teeth than
without sleeves. If the 3D preview indicates artifacts, check whether
the window of the disposable sleeve is clean and wipe it down with
alcohol.
Dentsply Sirona
66 37 479 D3492
D3492.201.07.28.02    09.2019