10 PC Diagnostic Tool

10.4 Troubleshooting

84
Operating Instructions for the Acquisition Unit - With CEREC Omnicam
➢ Check the result in the "Windows Stress Test" window.
Tip: You may have to rearrange the windows to be able to see the
"Windows Stress Test" window.
Test result: Pass
No errors were found on the PC-specific hardware. Replacing the PC
component or the PC is not advisable.
Test result: Fail
An error was found on the PC-specific hardware. Replacing the PC
component or the PC may be required.
An image restore is not advisable.
1. Check the "Windows Stress Test" window for those
components that did not pass the test.
2. Carry out the appropriate steps in the chapter entitled
Troubleshooting  [→ 84] .
Ending the "Sirona Windows Diagnostics" test
➢ Click the "Stop" button.
Ä The individual test windows close.
Ä The PC is switched off.
Troubleshooting
10.4
The recommended procedure is as follows. If troubleshooting and/or a
component replacement cannot eliminate the fault, replace the PC.
In this case, specify which test was defective when returning the
defective PC.
Dentsply Sirona
66 37 479 D3492
D3492.201.07.28.02    09.2019