Step-by-step instructions
3.1 Application
1. Insert Compules
tain that the collar on the Compules
2. Remove the colored cap from the Compules
360° to gain the proper angle of entrance into the cavity or to the mixing pad.
DO NOT USE EXCESSIVE FORCE.
Indirect use: to prevent Compules
fluids or contaminated hands it is mandatory that the Compules
side the dental unit with clean/disinfected gloves. Dispense the necessary amount
of material onto a mixing pad and protect against light. Apply material into the cavity
with a plastic instrument.
- or -
Direct intraoral use: dispense the material directly into the cavity preparation using
a slow, steady pressure.
3. To remove the used Compules
pulled back completely by allowing the handle to open to its widest position. Apply a
downward motion to the front end of the Compules
Excessive force.
Injury.
1. Apply slow and steady pressure on the Compules
2. Do not use excessive force. Compules
For further information please refer to composite Instruction for Use.
4
Hygiene
NOTICE
Wrong cleaning or disinfection method.
Damage to the Compules
1. Always disassemble Compules
2. Do not clean or disinfect with aggressive agents (e.g. solutions based on orange
oil or acetone).
3. Start reprocessing within 1 hour after use.
4.1 Disassembly and preparation for reprocessing
1. Partially close the Compules
hinge.
2. Push upwards and lift hinge. The Compules
and the plunger is exposed.
3. Clean Compules
4. Remove restorative material, if any, with a soft paper tissue and alcohol of 70% vol.

4.2 Manual cleaning and disinfection

1. Discard gloves according to local regulations.
2. Disinfect hands with an appropriate bactericidal, virucidal, and fungicidal hand disin-
fectant solution according to local regulations. Use according to disinfectant solution
manufacturer's Instructions for Use.
Tip into the notched opening of the Compules
®
®
Compules
Tips Gun may result.
®
Tips Gun.
®
Tips Gun and place your thumb on the rear part of the
®
Tips Gun using a soft brush and running cold tap water.
®
Tip is inserted first.
®
Tip. The Compules
®
Tip from exposure to spatter or spray of body
®
Tip, be sure that the Compules
®
Tips Gun before reprocessing.
®
Tips Gun is separated into two parts
®
Tips Gun. Be cer-
®
Tip may be rotated
®
Tip is handled off-
®
Tips Gun plunger is
®
Tip and remove.
Tips Gun.
®
Tip rupture or ejection from
®