Dentsply Sirona
Operating Instructions for the Acquisition Unit - With CEREC Omnicam
Monitoring window
66 37 479 D3492
D3492.201.07.28.02    09.2019
Explanation
When the battery charge is almost depleted, the shut-
down window opens. The operating system is then
shut down after 30s and the PC can then be switched
off.
A continuous signal is then sounded via the system
loudspeaker.
Fan blocked, status message in monitoring window.
Warning window with 30s countdown until the PC
shuts down.
A continuous signal is then sounded via the system
loudspeaker.
The temperature monitor has two message thresholds.
The first message threshold is output as the message
"High" in the temperature result field. The "High" read-
ing is displayed in the red-and-black flashing mode. No
countdown window appears since, depending on the
load and ambient conditions, the unit can keep operat-
ing for a few minutes, or even for a longer period of
time if the temperature level decreases again. Direct
shutdown occurs if the 2nd threshold is reached.
No battery is inserted.
Restarting delay
Once the power supply has been switched off, it can only be switched
back on again after 10 seconds have elapsed.