Dentsply Sirona
Operating Instructions for the Acquisition Unit - With CEREC Omnicam
66 37 479 D3492
D3492.201.07.28.02 09.2019
8.1.5.10
Dry heat sterilization of the mirror sleeve
NOTE
Dry heat sterilization and high-level disinfection must not be
combined.
The process for dry heat sterilization is as follows:
1. Pre-cleaning (see „Pre-cleaning of the mirror sleeve [→ 56]")
2. Remove the mirror sleeve from the CEREC Omnicam (see
„Removing the mirror sleeve [→ 55]").
3. Sterilize the mirror sleeve using dry heat at 190°C for 6 minutes
(unwrapped) or 12 minutes (wrapped). Use only pouches which are
suitable for dry heat sterilization cycle of at least 190⁰C, 12 minutes.
Alternatively, sterilize mirror sleeve using dry heat at 160°C for 120
minutes or for 60 minutes (unwrapped or wrapped). Use only
pouches which are suitable for dry heat sterilization cycle of at least
160⁰C, 120 minutes.
The CPAC Cox RAPIDHEAT Sterilizer, the CPAC Sterident Model
200 and the CPAC SteriSURE have been validated by Sirona
Dental Systems.
For 6 minutes 190°C program # 1 has to be set on the CPAC Cox
RAPIDHEAT Sterilizer.
For 12 minutes 190°C program # 3 has to be set on the CPAC Cox
RAPIDHEAT Sterilizer.
For 120 minutes 160°C the CPAC Sterident Model 200 is to be
used.
For 60 minutes 160°C the CPAC SteriSURE is to be used.
4. Store the mirror sleeve in such a way that it is protected from
contamination until the next use.
5. Carefully re-attach the mirror sleeve and allow it to lock in place
(see „Refitting the mirror sleeve [→ 55]").
NOTE
The mirror sleeve changes its color which will not have any negative
impact upon the mirror sleeve´s durability.
8.1 Care, cleaning, disinfection and sterilization
8 Maintenance