7 Operation
7.4 Directing the camera
48
Operating Instructions for the Acquisition Unit - With CEREC Omnicam
Completing the scanning process
1. Start as indicated above, on the occlusal surface of the right
terminal tooth, and scan it occlusally.
Tilt the camera by 45° in a palatinal direction (oral) and guide it from
the distal to the mesial.
2. Tilt the camera another 45° in a palatinal direction (oral) and move it
in a distal direction.
3. Tilt the camera by 90° on to the occlusal surface and move it in a
mesial direction.
4. Tilt the camera in a 45° buccal direction and move it back towards
the distal.
5. Then tilt the camera a further 45° in a buccal direction to a total of
90° and move it in a mesial direction again.
Dentsply Sirona
66 37 479 D3492
D3492.201.07.28.02    09.2019