5 Technical information
5.3 Electromagnetic compatibility
Interference immu-
IEC 60601-1-2 Test level
nity tests
Conducted RF in-
3 V
terference
150 kHz to 80 MHz
IEC 61000-4-6
Radiated RF inter-
3 V/m
ference
80 MHz to 800 MHz
IEC 61000-4-3
3 V/m
800 MHz to 2.5 GHz
26
Compliance level
3 V
eff
3 V/m
3 V/m
Remark 1
The higher frequency range applies at 80 MHz and 800 MHz.
Remark 2
These guidelines may not be applicable in all cases. The propagation of
electromagnetic waves is influenced by their absorption and reflection
by buildings, objects and persons.
1. Field strengths from fixed transmitters, such as base stations for
radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM/FM radio and TV broadcasts, cannot be predicted
theoretically with accuracy. An investigation of the location is
recommended to determine the electromagnetic environment
resulting from stationary RF transmitters. If the measured field
strength in the location in which the UNIT is used exceeds the
applicable RF compliance level specified above, the UNIT should
be observed to verify normal operation. If unusual performance
characteristics are observed, it may be necessary to take additional
measures such as reorientation or repositioning of the UNIT.
2. Over the frequency range 150kHz to 80 MHz, field strengths should
be less than 3 V/m.
Operating Instructions for the Acquisition Unit - With CEREC Omnicam
Electromagnetic environment –
guidelines
d= [1.2] √P
eff
d= [1.2] √P
at 80 MHz to 800 MHz
d= [2.3] √P
at 800 MHz to 2.5 MHz
where P is the nominal transmitter
output in watts (W) specified by the
transmitter manufacturer and d is
the recommended working clear-
ance in meters (m).
Field strengths from fixed RF trans-
mitters, as determined by an electro-
magnetic site survey
than the compliance level
frequency range.
Interference is possible in the vicinity
of equipment bearing the following
graphic symbol.
Dentsply Sirona
1
should be less
2
in each
66 37 479 D3492
D3492.201.07.28.02    09.2019