8 Maintenance

8.4 Replacing the main fuse

72
Operating Instructions for the Acquisition Unit - With CEREC Omnicam
Replacing the main fuse
8.4
WARNING
Electric shock
Disconnect the power plug at the unit end before replacing the fuses.
NOTE
Fuse type
Use only fuses of the same type in the fuse module!
A
Voltage selection insert
B
Main fuses
Fuses:
ü The power plug must be disconnected.
1. Unlatch the fuse module with a screwdriver and pull the module out.
2. Replace the defective fuses.
3. Reinsert the fuse module.
C
Fuse module
D
Window
T8A H 250V
Dentsply Sirona
A
B
C
D
Order No. 62 33 188
66 37 479 D3492