8 Maintenance
8.2 Calibrating CEREC Omnicam
66
Operating Instructions for the Acquisition Unit - With CEREC Omnicam
Calibrate the camera
1. Remove the protective cap from the calibration set.
2. Mount the calibration set on the tip of the camera until it locks into
place.
3. Secure the CEREC Omnicam in the calibration set using one hand.
Ensure that the external calibration set screw is fully screwed in a
clockwise motion until it gently locks into place.
4. Click on the "OK" button.
Ä The measuring process starts.
Ä The software prompts you to proceed to the next latching.
5. Turn the screw counter-clockwise until you reach the next latching
point.
6. Click on the "OK" button. In doing so, ensure that the CEREC
Omnicam does not move.
Ä The software confirms the calibration process.
Ä The software prompts you to proceed to the next latching.
7. Execute steps 5 and 6 a total of 11 times.
Ä The software provides status updates on the calibration and
informs you once the procedure is complete.
Ä You will be prompted to measure the position of the exit
window.