8 Maintenance
8.1 Care, cleaning, disinfection and sterilization
58
Operating Instructions for the Acquisition Unit - With CEREC Omnicam
NOTE
It is recommended that you never switch between disinfecting and
sterilizing solutions. However if you do, make sure you thoroughly
rinse the HLD set before switching.
4. Place the white protective cap on the camera tip and place the
camera in the camera cradle.
5. Use the following disinfectant for the high-level disinfection: CIDEX
®
OPA.
6. Slid the plug into the metal holder.
7. Press the mirror sleeve onto the plug while holding the metal holder
in order to ensure that no particle contamination or fluids penetrate
the inside of the mirror sleeve.
Ä If attached correctly, the plug seals the mirror sleeve watertight.
CAUTION
Observe the disinfectant's manufacturer's safety indications!
Dentsply Sirona
66 37 479 D3492
D3492.201.07.28.02    09.2019