8.1 Care, cleaning, disinfection and sterilization
Mirror sleeve for CEREC Omnicam
Pre-Cleaning process
(while mirror sleeve is attached to the camera)
1. Clean with a cotton gauze moistened with tap water of potable water quality that meets Federal Clean Water
Standards and pH neutral soap
or
use a clean lint-free cloth which has been soaked in CaviCide
or
use Cavi Wipes
or
use pdCARE Wipes
or
use a clean lint-free cloth which has been soaked in > 60% isopropyl alcohol
2. Dry the mirror sleeve with a clean lint-free cloth.
HLD process
®
1. Use CIDEX 
OPA as disinfectant for high-level
disinfection per manufacturer´s instructions.
2. Leave the mirror sleeve in CIDEX 
minutes at 20°  C .
3. Rinse the sleeve under tap water of potable water
quality that meets Federal Clean Water Standards.
4. Dry the mirror sleeve with a clean lint-free cloth.
56
Operating Instructions for the Acquisition Unit - With CEREC Omnicam
8.1.5.5

Mirror sleeve reprocessing overview

See detailed instructions after summary.
Dry heat sterilization process
➢ Dry heat 160°C (Sterident Model 200) for 120
®
OPA for 12
8.1.5.6

Pre-cleaning of the mirror sleeve

Cleaning process immediately after using the CEREC Omnicam: While
the mirror sleeve is attached to the camera, wipe off so that any surface
contamination cannot harden and adhere to the surface of the mirror
sleeve.
1. Use a cotton gauze moistened with tap water of potable water
quality that meets Federal Clean Water Standards and pH neutral
soap or use a clean lint-free cloth which has first been soaked in
one of the cleaning products listed in the „Cleaning and disinfection
agents  [→ 52]" section to remove visible contamination.
2. Wipe down the mirror sleeve afterwards with the absorbent cotton
gauze dipped in drinking water.
3. Next dry the mirror sleeve using a lint-free cloth.
minutes (wrapped or unwrapped)
or
Dry heat 160°C (SteriSURE) for 60  m inutes
(wrapped or unwrapped)
or
Dry heat 190°C (Cox RAPIDHEAT Sterilizer) for 6
minutes (unwrapped),
or
Dry heat 190°C (Cox RAPIDHEAT Sterilizer) for 12
minutes (wrapped).