5 Technical information

5.3 Electromagnetic compatibility

Emission measurement
RF emissions according to CISPR 11
RF emissions according to CISPR 11
Harmonics
according to IEC 61000-3-2
Voltage fluctuations / flicker according
to IEC 61000-3-3
24
Operating Instructions for the Acquisition Unit - With CEREC Omnicam
Dimensions and weight
Dimensions (WxHxD)
in mm
in inches
Weight
● without monitor and battery pack
approx.
● Monitor approx.
● Battery pack approx.
Electromagnetic compatibility
5.3
Observance of the following information is necessary to ensure safe
operation regarding EMC aspects.
CEREC AC complies with the requirements for electromagnetic
compatibility (EMC) according to IEC 60601-1-2:2001 and A1:2004.
CEREC AC is hereinafter referred to as "UNIT".
5.3.1

Electromagnetic emission

The UNIT is intended for operation in the electromagnetic environment
specified below.
The customer or user of the UNIT should make sure that it is used in
such an environment.
Conformity
Group 1
Class B
Class A
coincides
350 x 1210 x 470mm
13¾ x 47⅝ x 18½"
38 kg (83.8 lbs)
4 kg (8.8 lbs)
2 kg (4.4 lbs)
Electromagnetic environment - guidelines
The UNIT uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
The UNIT is intended for use in all facilities, in-
cluding residential areas and in any facilities con-
nected directly to a public power supply providing
electricity to buildings used for residential pur-
poses.