Step Cycle
3
Neutralizing
Rinse
4
Intermediate
Rinse
5
Disinfection
Rinse
6
Dry
After each cleaning cycle, inspect instruments for signs of cracking, crazing, or
degradation that may affect function.
Sterilization
Sterilize instruments before each use.
CAUTION: Sterilization temperatures must not exceed 138°C/280°F.
NOTE: Higher temperatures or longer sterilization times may be used; however;
increasing the sterilization time or temperature (not to exceed 138°C/280°F) of
the stated cycles may reduce instrument life.
Pre-Vac Sterilization Parameters
Temperature
Time

POSTOPERATIVE CONSIDERATIONS

Prophylactic antibiotic treatment should be considered for all patients undergoing
dental or other potentially bacteremic procedures.

Anticoagulation Therapy

Use of anticoagulants and antiplatelet agents may be prescribed according to
individual patient needs.
Carefully monitor the patient's anticoagulation status when postoperative therapy is
used. Surgeons who implant the SJM Séguin Annuloplasty Ring should be current
on anticoagulation regimes.

Disposal

The Instructions for Use is recyclable. Dispose of all packaging materials as ap-
propriate. Dispose of rings and accessories per standard solid biohazard waste
procedures.

LIMITED WARRANTY

The user must comply with the terms and conditions of this limited warranty in
order for this limited warranty to apply.
St. Jude Medical warrants that if the Product does not comply with the written
specifications for the Product, then St. Jude Medical will accept a return of the
Product for replacement for so long as the user requires the Product for use in
connection with the sizing or implantation of the St. Jude Medical™ ring product(s)
identified in the instructions for use, subject to the other terms of this Limited
Warranty including:
• The Product must be used in accordance with the instructions for its use and
maintenance.
• The Product must be returned for analysis to:
1 pH10.4 – 10.8 (2 – 10ml/l, determined in deionized water, 20°C). In tap water or
softened water as well as with carried-over pre-rinsing water, the pH value may
vary - Neodisher MediClean forte Technical Data Sheet, 05/2015.
Duration Cycle
Requirements
5 minutes
~25°C tap water
4 minutes
Cold tap water
5 minutes
93°C tap water
--
--
1
134°C/273°F
5 minutes
Comments
Drain when complete.
Drain when complete.
Drain when complete.
Allow to air dry
8