CALIbr = CALIBRATION ROUTINE
6.8
• To set the parameter press the [Func] key to show "CALIbr".
• Remove all weight from the platform.
• Press [Print] and the display will show "noCAL" while the scale is
measuring the initial mass (pan must be empty).
• Then the display will show "LOAd xxx" where xxx is the required
calibration weight.
• Place the weight on the platform then press [Print] – display will
show "CAL" and then (after calibration) "unLoAd".
• Remove the calibration weight – the scale will return to normal
weighing.
rePL = RESULTS PRINTED AUTOMATICALLY OR MANUALLY
6.9
• This parameter controls the operation of the RS-232 interface.
• The balance can be configured to either print automatically when
the balance becomes stable or only when [Print] is pressed.
• To set the parameter press the [Func] key to show "rePL".
• Press [Print] to see the parameter.
• Press [Print] to select either "rePL= 0" or "rePL= 1".
"rePL= 0"
"rePL= 1"
• Press the [Func] key to return to weighing.
© Adam Equipment Company 2005
Manual output when [Print] key is pressed.
Automatically print when the results are stable.
13