"No Response" on the Display
When the Remote Control displays "No Response," there is a communication problem between the Remote Control and the
P
Stimulator. Press the
button to activate automatic searching for the Stimulator. Then reposition the Remote Control until it beeps.
This indicates that communication has been established.
Once the Remote Control connects with your Stimulator, you will be returned to the display you were using before the problem began.
If the problem persists, recharge the stimulator and check to see if the problem is solved.
Occasionally, telemetry problems occur because orientation or interference prevent the Remote Control from finding the Stimulator.
P
Move the Remote Control closer to your Stimulator and then press the
button.
Call your healthcare professional if the problem continues.
Patient System Handbook
90868426-08 REV A 38 of 213