Fault description
Fault description
no display
TACHO - ERROR 01, 02
IMBALANCE / UNWUCHT
CONTROL - ERROR 04,
06-09
N > MAX 05
AB5005en
The use of certain accessories is time-limited. For safety reasons, the
accessories must no longer be used when either the maximum number
of permissible run cycles marked on them or the expiry date marked on
them has been reached.
■
The maximum permissible number of run cycles or the expiry date can
be seen on the accessories.
■
The centrifuge is equipped with a cycle counter.
CAUTION
Risk of injury from broken glass.
Broken glass may cause glass splinters and contaminated
liquids to be found inside the centrifuge.
−
Wear cut-resistant gloves.
−
Wear protective goggles and a face mask.
Broken parts of the tube, glass splinters and spilled centrifuge material
must be removed completely in the event of leakage or if a centrifuge
tube breaks. Glass splinters that are not removed will cause further glass
breakage.
The rubber inserts and the plastic sleeves of the rotors must be replaced
after a glass breakage.
Disinfection must be carried out if the material is infectious.
Customer service must be notified if the fault cannot be rectified based on
the fault table. State the centrifuge type and serial number. Both numbers
can be seen on the type plate of the centrifuge.
* Error number does not appear on the display.
Cause
No power. Triggering of the overcur-
rent protection fuse (for centrifuges
5005-08 only).
Tacho defective. Motor, inverter,
electronics defective.
The rotor is unevenly loaded.
Lid lock error.
Overspeed error
Rev.: 20 / 01.2024
Troubleshooting
Remedy
■
Check the supply voltage.
■
The mains switch is in switch
[I]
position
■
Open the lid.
■
Set the mains switch to
■
Wait at least 10 seconds.
■
Turn the rotor vigorously by hand.
■
Set the mains switch to
rotor must rotate while switching
on.
■
Open the lid or hatch.
■
Check the loading of the rotor.
■
Repeat the centrifugation run.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
[0] .
[I] . The
45 / 50